10. Trouble Shooting
Q1 Power is not ON
‧ Check if the plug is connected to mains.
‧ Check for a blown fuse.
Q2 Low Pressure Alarm is on
‧ Check if the Quick Connector is tightly secured.
‧ Check if all tubing connections along mattress are secured.
‧ Check if the air hoses are kinked or broken.
‧ Check if the CPR knob is tightly closed.
Q3 Power Failure Alarm is on
‧ Check if the power is suddenly shut down.
‧ Check if the power cord is connected properly.
Q4 Patient is bottoming out
‧ Pressure setting might be inadequate for the patient, increase comfort range 1 level higher and wait
for a few minutes for best comfort.
Q5 Mattress form is loose
‧ Check if all the snap buttons or straps of mattress are all securely fastened.
‧ Check if the mattress is fixed to the bed frame by straps.
Q6 No air produced from some air outlets of the air tube connector
‧ This is normal since there is alternating mode. Air outlets take turns to produce air during their
cycle time.
Q7 System malfunction (Code 11 or 12)
‧ Check if there is only one or two knobs at the foot-end of the mattress remained open.
If the above information does not solve your problems, please contact your local agent directly. They
might require a technician to take care of the problem.