Data Extraction; Printing The Data Stored In The Lrf-04 - CESVA LRF-04 Manual Del Usuario

Limitador registrador frecuencial
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

13/09/99
14/09/99
22:30 <--> 03:00
Emi.day lim.:
90 dBA
Rec.day lim.:
40 dBA
Disconnection:
0 s
Sensor inac.: 0 s
LF max :
87.0 dBA
Exceed.: 1235 >85 dBA
-
BATTERY RUN OUT: Before the battery runs out, the LRF-04 records the day and time
when this occurred. You may consult the last 10 times this happened by selecting this
option from the display menu.
-
REPROGRAMMING: This option allows you to consult the last 10 dates on which the
LRF-04 was programmed.
DELETIONS: This option tells you the last 10 occasions on which the LRF-04 data was
-
deleted.
-
TOTAL COUNTERS: This option allows you to consult the total counters, which are the
following: emission exceeded (in seconds), reception exceeded (in seconds), sensor
inactive (in seconds) and total number of deletions.
-
VERSION AND SERIAL NUMBERS: Showing the version and serial numbers of the
LRF-04.

5. DATA EXTRACTION

As we saw in the previous chapter, all the data stored in the LRF-04 can be consulted on
screen from the keyboard. This information may also be printed out by a parallel printer or
transferred to a PC via serial connection or modem

5.1 Printing the data stored in the LRF-04

With the numerical keyboard and the LCD display it is possible to print the programming and
all the data stored in the LRF-04 as follows:
Connect a printer to the PRINTER output of the LRF-04.
Press PRINT (F) on the numerical keyboard. The printing menu will appear on the LCD
screen.
To move around the printing menu use the arrow keys (A) and (B). To select an option,
press ENTER (C) and to quit an option press ESCAPE (E).
While printing is in progress, the LRF-04 continues to function, and after two minutes
without any key being pressed the LRF-04 automatically quits the printing menu.
The printing menu has the following options:
Programming: This allows you to print a report containing all the parameters
-
programmed in the LRF-04.
Night timetable:
22:00 <-->
Emi.nig.lim.:
85 dBA
Rec.nig.lim.:
35 dBA
MaxLeq1':
86.4 dBA
14/09/99
02:55
MinLeq1':
UND dBA
13/09/99
23:51
.
34
Leq :
08:00
Attenuat. time:
Emissi. exceed: 0
Recep. exceed: 0
Percentiles (dBA)
L1
: 86.3
L5
: 84.2
L10 : 81.5
83.2 dBA
2 s
L50: 81.1
L90: UND

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Enos lrf-04

Tabla de contenido