Note:
- The current measurement should be taken within 0°C~40°C. Do not suddenly release the
trigger, as the impact may change the reading for a short time.
- To ensure measurement accuracy, center the conductor within the jaws. Otherwise, ±1.0%
additional error in reading may be produced.
- When testing high current, the clamp will vibrate slightly, which is a normal phenomenon.
DC CURRENT MEASUREMENT (PICTURE 1)
1) Turn the function dial to the
2) Short press the SELECT button to switch to DC current measurement. If the display on the
LCD is not zero, short press the REL ZERO button to enable zero clearing.
3) Press the trigger to open the clamp jaws, and fully enclose one conductor (only one
conductor can be measured at a time). For optimum results, center the conductor within the
jaws.
Note:
- The current measurement must be taken within 0°C~40°C. For DC current measurement, if
the reading is positive, the direction of current is from top to bottom (from panel to cover). Do
not suddenly release the trigger, as the impact will change the reading for a short time.
- To ensure measurement accuracy, center the conductor in within jaws. Otherwise, ±1.0%
additional error in reading will be produced.
- After DC current (especially large current) measurement, the open circuit base may be too
large. Please do an AC current test to eliminate the residual magnetic signal generated by
the jaws.