Descargar Imprimir esta página

Steinberg Systems SBS-ZW 3001H Manual De Instrucciones página 6

Balanza de peso
Ocultar thumbs Ver también para SBS-ZW 3001H:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 18
WEIGHT CALIBRATION
During self-checking, press
checking, the three windows will display:
Use
button to select the unit (Kg and g available), input "090316" by using the
number keys and then press
acity selection", three windows display as:
1.
"Cap
(xxx mean the capacity) input the capacity by using use number keys (example:
30Kg scale, input 30. If unit is g, then input 30000) press
2.
"Division selection", three windows display as
(x.x means the division). Press
1/600000 available), press
3.
"Zero track range selection", the three windows show as:
button (there are 0.5, 1.5, 2.5, 3.5 available, bigger value means that the zero-
track range will bigger). Press
EN
4.
Calibration mode selection", press
There are:
A:
"Calibration by any weight".
B:
"Calibration by two points".
C:
"Calibration by three points".
Operation as below:
A.
confirm, display will show the capacity value. At this time input the weight value by
keyboard, then put the weight same just input that weight value on the pan. Press
button
when display is stable and on the weight windows show the number same
the weight value. The calibration is finished.
NOTE: when the calibration by any weight is selected, the weight value couldn't be
smaller than 1/3 of full capacity.
B.
for calibration by two points. Press
code will be shown, press
button and display will show "HALF", put on the pan 1/2
of full capacity, wait until the number will be stable. Press
show "FULL". Put on the pan full of capacity and wait until it will be stable. After press
button to confirm. The scale will display the weight value and the calibration is finished.
and
buttons at the same time. After the self-
.
button to enter settings.
button to confirm.
.
button to choose the division (there are 1/3000-
button to confirm.
to confirm.
button for calibration mode selection.
for calibration by any weight, press
button, AD inner
button again, display will
10
C.
Press
show "LOAD1", put on the pan 1/3 of full capacity weight. Press
display will show "LOAD2", this time put on the pan 2/3 of full capacity weight, press
button to confirm. When display shows "LOAD3", put on the pan full capacity of
weight, press
calibration is finished.
NOTE: during the set, press
Simple calibration:
During any weighing, press
show as:
, press
Input calibration weight value by number keys and put it on the weight, wait until the
weigh is stable and then press
REGULAR TROUBLESHOOTING
1. Unstable weighing
Low voltage for accumulator: Switch off the scale immediately and recharge it
more than 12 hours.
Low intensity or strong shake for platform: Strengthen the platform or avoidshake.
Influenced by the wind: Set a glass cover to reduce the influence of the wind.
2. No display when switch on the scale
Accumulator is unusable: Change the accumulator if it remains invalid after
button to
recharging.
Broken with fuse or switch: Change the fuse or switch.
Others: Send it for maintain.
TRANSPORTATION AND STORAGE
Avoid shock and dampness during transportation.
Avoid strong sunshine and rain during storage.
ATTENTION
1.
Please read this instruction manual carefully before you use the scale.
2.
If it has not be used for long time, please recharge it once 2 months at least,
andkeep the power switch on "off" status to avoid discharge of accumulator.
3.
LCD of the scale must to be operated under rich light, you are advised to select
product with backlight under low light condition.
Rev. 22.12.2021
Rev. 22.12.2021
For calibration three points.
button, display will show AD inner code. Press
button to confirm. The scale will display the weight value and the
button to go back to up menu.
button for about 3seconds, that three windows will
button to confirm. The calibration is finished.
11
button again, display will
button to confirm,
EN

Publicidad

loading