SOFTTOUCH FACE MASKS
Dispersion
Limiting Vent
During the inspiratory phase,
the tab, acting as a Dispersion
Limiting Device, bends inwards
towards the mask.
CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from
the socket.
DEVICE (A) AND TUBING EXTERIOR (B)
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of
any sort).
ACCESSORIES
Open the nebuliser by turning the upper part (C1.2) anticlockwise, remove the upper
part (C1.2) as shown in the "Assembly diagram" in point C1.2.
Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one
of the methods described below.
method A: Sanitise accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C3-C4-C5 uunder
potable hot water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C3-C4-C5
dishwasher with a hot cycle.
method C: Sanitise accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C3-C4-C5 by
immersing them in a solution of 50% water and 50% white vinegar, then rinse
thoroughly under potable hot water (approximately 40°C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of
the methods described below.
method A: Accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C3-C4-C5 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of
potable water and disinfectant, according to the proportions indicated on the
packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of
Soft
biocompatible
material
SoftTouch masks have an outer edge made of soft
biocompatible material that ensures excellent
adherence to the face, and is also equipped with
an innovative Dispersion Limiting Device. These
distinctive elements that distinguish it allow great-
er sedimentation of medication in the patient and
also limit dispersion.
12
During the expiratory phase, the
tab, acting as a Dispersion Lim-
iting Device, bends outwards
from the mask.