Sample Rate:
2 samples/sec for the digital display
20 samples/sec for the analog bargraph
Power Supply: 9V, NEDA 1604 (6LR61) alkaline battery
Low Battery Indication:
is below the required voltage
Battery Life: 40 hours approx
Auto-Power Off: The meter will power itself OFF if there
is no push button or rotary switch operation for 30 minutes.
To deactivate this function, press the "MAX/MIN" button
and keep it pressed down, then power up the probe.
Jaw opening size: 1.575" (40mm)
Dimension: 9.53 x 2.60 x 1.42" (242 x 66 x 36mm)
Weight: 14.10 oz (400g) with batteries
3.3
Environmental Specifications
Altitude: 2000 meters
Operating Temperature:
-14° to 122°F (-25 to 50°C), 80% RH, non-condensing
NOTE: If the meter is to be used below 32°F (0°C), we suggest
that the battery be replaced to ensure proper results.
Storage Temperature:
-14° to 140°F (-25° to 60°C) < 80% RH, battery removed
3.4
Safety Specifications
EN 61010, 300V, CAT IV, Pollution Degree: 2
EN 61010, 600V, CAT III, Pollution Degree: 2
EN 61010, 1000V, CAT II, Pollution Degree: 2
10
displayed when the battery
Clamp-on Meter Model 514