Connect System Components - Testomat EVO TH Serie Manual De Instrucciones

Dispositivo automático de análisis en línea para dureza total residual (dureza del agua)
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Terminal desig-
nation
Mass/PE
N (U)
L (V)
n
l
Connection example
Limit value contact LV 1
switches mains voltage
!
DANGER
Type
Function
IN
Mains protective conductor (5x)
Grid, N=Neutral conductor
(U=24V)
IN
Grid, L=Phase (V=24V)
Neutral conductor, secured with
4A (4x)
OUT
Phase, secured with 4A (4x)

Connect system components

Loosen the strain relief of the cable ducting (union nut).
Take out the blanking plugs of the corresponding cable ductings.
Push through the cable of the component.
Tighten the union nut of the cable ducting and so establish the
strain relief.
Connect the system components to the output terminals of relays 1
to 4 (e.g. valves)
If the system components require mains power, connect the
switched mains voltage (l) to the root contact
relay (see accompanying connection example for 230 VAC)
Connect the neutral conductor of the system component with one
of the terminals (n)
For components with a protective conductor connection, connect
them to the PE connection.
Ensure that the cores in the terminals are securely in place
(Illustrated position of relays: Device without current, mains: 230 V)
Indicate external voltage on the relay contacts!
If you connect system components that do not operate using the de-
vice voltage, you can apply external voltages to the relay contacts.
This external voltage cannot be switched off via the device mains
switch
There is a danger of electric shock!
Affix a warning to the device in this case (e.g. a sticker as shown left).
Assembly
Note
Only for mains supply of
230 V and 100 – 240V!
Power input 50-60 Hz
24 V / 100-240 V / 230 V
Grid for consumers, max. 4
A
of the respective
21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Evo th 2005Evo th 2025Evo th 2100Evo th 2250

Tabla de contenido