Water Shortage; Error History - Testomat EVO TH Serie Manual De Instrucciones

Dispositivo automático de análisis en línea para dureza total residual (dureza del agua)
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

WATERLOW COUNT
Count
SD Card n. inserted
12.09.13 09:00
12.09.13 09:00
12.09.13 09:05

Water shortage

The water shortage error is a special case, since one further setting
can be configured for this error. In the menu, under
Basic program
=>
the number of consecutively occurring errors before the alarm is trig-
gered can be set. An error is then generated for each measurement
cycle.
Number of water shortage errors
0
This function is specially provided for systems subject to intermittent
low water pressure errors, which trigger a "
device. The number 0 can be set for an immediate message, or the
setting can be raised to 250 occurrences before the error alert is
sent.

Error history

Service
Select =>
to call up error history.
Show(OK)
Select "
Confirm the input with "OK"
Select an error message from the list using the arrow keys.
The time and date of the occurrence are shown.
Again, press "OK".
Now –
depending on the error type
are shown.
If the error remains current, no end time is registered.
To delete the entire list, select
Delete(OK)
The number indicates the number of stored notifications.
The error history has room for up to 20 notifications.
Input basic programming data
Waterlow count
=>
Water low
Error History
=>
", to show the list of error messages.
– the start and end of the error
20
" error in the
45