Getting started
Once you have completed the steps in the section
tor, you can switch on the device.
Automatic venting
NOTE
After the device is switched on the indicator lines are automatically
vented and the measuring chamber purged, until indicator is detected
in the measuring chamber. The automatic venting cannot be inter-
rupted. Wait until the venting process has ended and acknowledge
the "voltage failure" error message by pressing the horn key. The
device is then ready for use.
You can skip automatic venting by keeping the OK button pressed
when switching on.
Since no settings in the programming menu can be configured while
the analysis is underway, press the PAUSE button once the venting
is complete or proceed to the programming menu before the first
analysis starts!
Once the first analysis has been successfully completed, you can
adapt the device to your desired usage. The following sections will
set out all the configuration options.
Now make the following settings, since these are imperative for
obtaining usable measurement values:
- Select
Indicator type and container size
- Select display
unit of measurement values
Now perform the first measurement by pressing the button. Upon
completion of the analysis, a measurement value must be shown.
If any error occurs, please refer to the section on
es/troubleshooting
and resolve the error.