FEATURES
The GIMA forehead thermometer was designed with advanced infra-red technologies and room temperature
compensation technology for immediate and accurate measurements.
Do not use this device in situations of vital importance or for applications concerning safety.
Specifications
Temperature measurement range: Forehead mode: 34~42.2°C (93.2~108°F), Scan mode: -22~80°C (-7.6~176°F)
Operating temperature range: 10~40°C (50~104°F)
Storage temperature range: -20~50°C (-4~122°F), RH 85%
Transportation temperature shall be less than 70°C, RH 95%
Accuracy: For Forehead mode: +/-0.2°C (0.4°F) during 36~39°C (96.8~102°F),+/-0.3°C (0.5°F) for other range.
For scan mode: +/-0.3°C (0.5°F) during 22~42.2°C (71.6~108°F), for other range +/-4% or +/-2°C (4°F)
whichever is greater.
Battery: one lithium battery (CR2032)
FUNCTION
Before each use make sure the probe is clean and without damage.
Press ON/MEM button. The thermometer is ready for use after you hear two beep sounds.
Setting date and time
Correctly setting the clock will assure a proper interpretation of the stored
measurements.
Press and hold the ON/MEM button for 5 seconds. When the Set icon starts to flash, it indicates that you
are in the setting screen. Use the ON/MEM button to scroll the 12/24 hour settings, the year, month and
date. Use the SCAN button to set the clock.
Choosing between Fahrenheit and Celsius
To change the unit of measure of the temperature between °C and °F, press and hold the
SCAN button with device switched off, then press and release the ON/MEM button until the
display shows the symbol °F or °C. Repeat the procedure to change the unit of measure again.
Note: holding the thermometer too long may cause a higher ambient temperature
reading of the probe. This could make the body temperature measurement
lower than usual.
Forehead temperature measuring
Make sure the probe lens and the forehead are clean and dry.
1) Press the ON/MEM button. After you hear two beep sounds and the icon
appears, the thermometer is ready for use.
2) Approach the thermometer to your temple, analyzing the area shown
in the figure.
3) Press the SCAN button and detect the value at close range around the
temple, with a very slow twisting motion.. While scanning, you will hear a
beep sound, which indicates the newest measurement is taking place.
Measurement has been completed after two short beeps are heard, after
5-30 seconds from the beginning of the measurement. The temperature
is now shown on the display. This is just an indication of the temperature
and determines only whether there is a temperature or not.