The light source: super high pressure mercury lamp which provides strong enough exciting light
(shorter wavelength light) for excitation of fluorescent specimen.
Exciting filter: only pass the excitation light of corresponding specimen, the rest of light is cut
off.
Fluorescent specimen: the specimen dyed with the fluorescent pigment.
Cut filter: blocks the excitation lights through the specimen and make observation easier.
Figure 2: Diagram of epi-fluorescence
2. BEFORE USE THE EPI-FLUORESCENCE MICROSCOPE
a) Place of installation
o Microscope must be placed on a firm and flat desktop or working table.
o Microscope must not be placed under direct sunlight and also keep it away from the
wall 20 cm.
o When connecting microscope with power supply, turn the main switch to "off"
position, make sure the inlet voltage must be in compliance with label marked on
the microscope, then plug into the electrical outlet. Please pay attention to make
good earth connection; otherwise, it would influence the electrical safety and
instrument performance.
o Please cover the microscope with dustcover when it is not in use
o Operating environment: the highest elevation is 2 meters. The ambient temperature
is 5-35ºC. the relative humidity <80%
b) Replacing the light bulbs
o When replacing the light bulbs, shut off the main switch and pull the power plug out
from the electrical socket; never pull power cord directly with hand by force.
o The bulb in light cabin can be replaced after it is cool.