Installation Of Halogen Lamps - Auxilab HBF001 Manual De Instrucciones

Microscopio de epifluorescencia
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1. Release the lamp house locking screw
2. Insert into lamp house as the direction of the arrow
3. Lock and tighten the locking screw
b) Installation of halogen lamps
The down illumination light source of this microscope is 12V/50W bulb.
Warning!!
Whenever replacing the bulb, first turn off the main switch and wait for complete cool of light
bulb, the lamp socket and house.
Steps:
1. Release the locking screw (left side of the box)
2. Remove the lamp house side cover
3. Wear gloves or a piece of gauze to hold bulbs; insert the bulb pins into lamp holders and
make sure the pins are inserted completely.
Do not touch bulb with our fngers; if occasionally fingerprint is left on the light bulb, use
a soft cloth to clean
4. Install the side cover of the lamp house
5. Tighten the locking screw
Manual de instrucciones- Instruction´s manual