Coltene COMPONEER Instrucciones De Uso página 19

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

(5) placing Componeer in end position
While holding the COMPONEER in position remove obvious excess and shape the composite to the COM-
PONEER. A matrix previously placed in the interdental space interferes with adhesion of the teeth. The
light-curing cannot be started until the correct position of the veneer has been verified. The patient should
be relaxed and facing straight ahead for the check. Defects that could extend to the palate can be filled
with dentine mass from the palatal direction beforehand.
(6) light-curing
The light-curing with the polymerisation lamp is best started from the palatal side to prevent build-up of
excessive stress. The palatal light-curing is continued for 40 s at less than 1000 mW/cm2. Then the maxi-
mum power is applied from the facial side for 40 s. If a polymerisation lamp with an output less than 1000
mW/cm2 is not available, it can be placed at a greatest distance (approx. 5 mm).
After curing and before trimming check that composite is in all the required places, otherwise corrections
must be made at this point. Major corrections can be made with SYNERGY D6, SYNERGY D6 Connect com-
posite, smaller fine corrections can be made with SYNERGY D6 Flow.
19

Publicidad

Tabla de contenido
loading

Tabla de contenido