4.6.3.2 Sample weight and volume (sample quantity)

Fig. 71
Fig. 72
The Sample Quantity (W) item is used to select whether one is wishing to use a sample weight or a
sample volume for titration or solution preparation.
You have the following options:
Manual sample weight: The sample weight is enquired by a prompt at the start of the method
and manually input.
Automatic sample weight: The sample weight is automatically transferred by a connected
balance.
Fixed sample weight: A fixed sample weight is input in g. This weight will then automatically
be used for each start of the method.
Manual sample volume: The sample volume in ml is prompted at the start of the method and
manually input.
Fixed sample volume: A fixed sample volume is input in ml. This volume will then
automatically be used for each test of the method.