Xylem SI Analytics TitroLine 5000 Manual De Instrucciones página 122

Ocultar thumbs Ver también para SI Analytics TitroLine 5000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

122
<Measuring speed> or drift will determine the span of time after which the measured value will be
accepted following a titration step.
Fig. 89
Drift-controlled acceptance of the measured value in terms of mV/min is set by selecting "normal",
"fast or "user-defined". The drift values at predefined in terms of in mV/min for normal and fast drift:
Normal drift
20 mV/min
Fast drift
50 mV/min
Small drift value = slow and precise
Large drift value = fast and "less precise"
The following parameter selection can be made for user-defined drift setting:
Minimum holding time [s]:
Maximum holding time [s]:
Measuring time t: [s]
Drift [mv/min]
Fig. 90
If normal or fast drift was selected before, the values will be defaulted for user-defined drift. In the
present case, for instance, 20 mV for normal drift:
01 - 99
01 - 99
01 - 99
01 - 99

Publicidad

Tabla de contenido
loading

Tabla de contenido