Sample Weight And Volume (Sample Quantity) - Xylem SI Analytics TITRONIC 500 Manual De Instrucciones

Ocultar thumbs Ver también para SI Analytics TITRONIC 500:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

90
Meaning of the individual factors:
W:
Weight of the sample in g
Fa:
Soluble foreign-matters portion in %
Fb:
insoluble foreign-matter portion in %
Fc:
Conversion factor for it unit
g/l
mg/l und ppm
g/100 ml
%
Fd:
Target concentration of the solution to be prepared in g/l, mg/l (ppm), g/100 ml, or %
Fe:
Specific weight of the weighted-in sample in g/cm³
Ff:
Volume correction in ml. this volume correction is the required surplus dosage for compensating the
volume contraction and the specific-weight difference between the sample weight and the solvent (please
observe the note on volume correction)
Fg:
Specific weight of the solvent used in g/cm³
Note on volume correction:
The user has to decide on a case-by-case basis whether a volume correction is necessary and according to
which procedure this correction is to be performed. As a rule, this volume correction may be omitted in the case
of solutions with very low percentages of diluted substance.

4.5.2.3 Sample weight and volume (sample quantity)

Fig. 61
Fig. 62
Chapter 4 Method Parameters
= 10
= 10000
= 1
= 1

Publicidad

Tabla de contenido
loading