Guidelines and manufacturer's declaration - Electromagnetic
resistance
The mattress is intended for operations in surroundings described below.
The customer or the user of the mattress needs to make sure that it will be
operated in surroundings of this type.
Resistance
IEC 60601
tests
Test level
± 6 kV
Discharge of
contact
static
discharge
electricity (ESD)
pursuant to
± 8 kV air
IEC 61000-4-2
discharge
± 2 kV for
power
Quick stray
cables
electrical
factors / bursts
± 1 kV for
pursuant to
input and
IEC 61000-4-4
output
cables
± 1 kV
Series
mode
voltage
Surges
pursuant to
IEC 61000-4-5
± 2 kV
Common-
mode
voltage
< 5 % VT
(>95 %
drop in VT )
for ½ period
40 % VT
(60 % drop
Voltage drops,
in VT )
Short-term
for 5
disruptions
periods
And
Fluctuations in
70 % VT
the supply
(30 % drop
voltage
in VT )
pursuant to
for 25
IEC 61000-4-11
periods
< 5 % VT
(>95 %
drop in VT )
for 5
seconds
Magnetic field
for the
Supply
frequency
3 A/m
(50/60 Hz)
pursuant to
IEC 61000-4-8
NOTE VT is the grid alternating voltage before applying the test level.
Compliance
Electromagnetic
level
surroundings -
Guidelines
Flooring needs to be
made of wood or
concrete
± 6 kV contact
or covered with
discharge
ceramic tiles
If the floor is covered
± 8 kV air
with synthetic
discharge
material, then relative
humidity needs to be
at least 30%.
± 2 kV for
The quality of the
power cables
power voltage needs
to meet standards
± 1 kV for input
typical in a
and output
commercial or
cables
hospital environment.
± 1 kV Series
The quality of the
mode voltage
power voltage needs
to meet standards
± 2 kV
typical in a
Common-
commercial or
mode voltage
hospital environment.
< 5 % VT
The quality of the
(>95 % drop in
power voltage needs
VT )
to meet standards
for ½ period
typical in a
commercial or
40 % VT
hospital environment.
(60 % drop in
Should the user of
VT )
the mattress require
for 5 periods
the continuous
function even with
70 % VT
the occurrence of
(30 % drop in
disruptions to the
VT )
power supply, it is
for 25 periods
recommended that
the nursing bed
< 5 % VT
receive undisrupted
(>95 % drop in
power from an
VT )
emergency source or
for 5 seconds
a battery.
Magnetic fields with
the grid frequency
should comply to
those normal in the
3 A/m
commercial or
hospital
environments.
Guidelines and manufacturer's declaration - Electromagnetic
The mattress is intended for operations in surroundings described below.
The customer or the user of the mattress needs to make sure that it will be
operated in surroundings of this type.
Immunity test
IEC
60601
Test level
3 Vrms
Conducted RF
150 kHz
IEC 61000-4-6
to 80 MHz
3 V/m
Radiated RF
80 Mhz to
IEC 61000-4-3
2,5 GHz
NOTE 1 At 80 MHz and 800 Mhz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular / cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment
due to fixed strength in the location in which the mattress is used exceeds
the applicable RF compliance level above, the mattress should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating
the mattress.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be
less than 3V/m.
5
MAT-X1, MAT-X2, MAT-X3
2018-07
resistance
Compliance
Electromagnetic
level
surroundings -
Guidelines
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the mattress, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
3 Vrms = V1
3 V/m = E1
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
a
survey
, should be less
than the compliance level
in each frequency range
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
b
.

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Mat-x2Mat-x3

Tabla de contenido