Xylem Si Analytics ViscoClock Manual De Instrucciones página 27

Aparato de medicón de la viscosidad
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.3 Placing the viscometer into the Viscosity Measuring Unit ViscoClock
Press the fixing plate (10) upwards against the handle plate (12), → Exhibit 3.
The following applies to Ubbelohde viscometers:
Insert the viscometer from below in an angle through the guidance plate (3), and then insert it
vertically into the centering prism (5).
Lower the fixing plate (10).
Check proper seating of the silicone plug on filling tube (3) (silicone plug 11D) and capillary tube
(2) (silicone plug 11S).
The viscometer is fastened, → Fig. 4.
The following applies to Micro Oswald viscometers:
Insert the Micro Oswald viscometer into the stand (6) of the Viscosity Measuring Unit ViscoClock
in such a manner that the capillary tube is located as is shown in Exhibits 3 + 4, but the filling
tube is inserted into the oblong hole of the guidance plate {9).in this positlon the filling tube of the
viscometer is located under the sillcone membrane (17).The membrane rod (16) is located in its
upper position - filling tube is open.
Lower the fixing plate (10).
The hand pump "sucking" VZ 6554 is a necessary prerequisite for Micro Ostwald viscometers.
Fig. 3
Initial operation
Fig.4
25

Publicidad

Tabla de contenido
loading

Tabla de contenido