4. CLEANING AND MAINTENANCE

Do not use abrasives substances or solvents. The device can be damaged.
DATOSPIR
spirometer requires, like other electromedical equipment, maintenance
aimed at assuring patient safety as well as operator and environment safety; and at ensuring
reliability and accuracy of the functions for which it has been developed.

4.1 SPIROMETER CLEANING

Clean the spirometer with a cloth moistened with water and neutral soap or with 96º alcohol.
Dry the remaining moisture. Make sure no liquid or foreign material enters the equipment
nor connectors or connections, especially in the pressure ports of the disposable transducer
handle (wipe the handle downwards).

4.2 TRANSDUCERS DISINFECTION

The transducer is the main part exposed directly to the patient. Therefore, it is necessary
to keep it in perfect physical and hygienic conditions.
To disinfect the transducer, proceed as follows:
A) TRANSDUCER FLEISCH
1.
Remove the filter and then the transducer by slightly
pressing the tab to release it from its housing.
2.
Immerse the transducer and filter in a CIDEX® OPA
disinfectant solution (see manufacturer's instructions).
Rinse with distilled water.
3.
Shake the transducer to remove residual water, dry it at room temperature and
assemble again.
EN 18
DATOSPIR
· User's Manual
511-D00-MUM · Rev 1.01