DO NOT disinfect the disposable accessories. The use of disinfectant products
can affect the accessories integrity inducing a loss of accuracy in the measurement.
Lilly disposable transducer does not need any cleaning. It is for single use. It must be
disposed once used by the patient.
4.3 PREVENTIVE MAINTENANCE
The preventive maintenance include those actions aimed at keeping the device in good
condition of use.
Actions to be performed by the user:
Perform an internal review of the device periodically by accessing the DEVICE
1.
MAINTENANCE / CHECK menu (this operation can be performed through the W20s
Software).
2.
Check that the connections, accessories and external elements of the device are in
good condition for use and there are no damages. Pay special attention to the cable
and USB connector.
3.
Check the calibration daily and calibrate when required (this operation can be performed
through the W20s Software).
4.
Define the period between maintenance and calibrations in the CUSTOMIZATION MENU
of the device. The device will warn every time it is turned ON if the period is exceed.
If "0 days" is set, there will be no warning (this operation can be performed through
the W20s Software).
If there is any anomaly that cannot be solved, contact
your dealer.
EN 20
DATOSPIR
SIBEL S.A.U.
After Sales Service or
511-D00-MUM · Rev 1.01
· User's Manual