en - Areas of application
A 806 holder for
Single-Site® in-
struments with
rinse connection
on the shaft
48
Push the hose adapter into the rinse connection on the shaft of the
instrument.
The clamp is used to secure the hose adapter.
If the hose adapter has come loose from the rinse connection on
the instrument during reprocessing, it must be reprocessed again.
Fit the instrument into the holder from below.
The pipe connection on the holder must be pushed into the rinse at-
tachment on the instrument.