Parts Counting
Two types of counting modes are available, standard or enhanced.
N
N
1
N
1
Exiting the .U.N.I.T. Menu
Use the same procedure as Exiting the .S.E.T.U.P. Menu.
Calibration
Span Calibration
Span calibration uses two calibration points, zero and a specified calibration weight.
Before beginning calibration, make sure the Lock Switch is off. Clear the platform.
NOTE: Value of calibration mass depends on capacity of balance. After
calibration, the balance returns to the currently selected weigh mode.
Start with the balance OFF.
5
• • •
Y
16
2014_391_10889_4_SPU_Portable_Series.indb 16
1
OR
(See page 17 for details)
(Release button)
(Acquiring zero value)
1
www.eu.fishersci.com/fisherbrand
Y
1
N
1
Y
1
(Place mass value on platform)
(Example)
23/01/2015 11:37:34