TROUBLESHOOTING

Please find solutions from the following table first in case of any problem
during operation.
NO. Faults
1
Lower
nebulizing rate.
2
Power LED
does not light
up and the
device does
not work.
3
Power LED
lights up but
the device
does not work.
If faults still exist after take measures as listed above, contact your supplier
or your nearest local distributor.
ENGLISH
Symptom/Possible Causes
Low battery indicate LED
lights up/goes out.
Nebulizer is foul.
Battery's polarity has
been reversed.
Low battery voltage
Power adapter is not
correctly connected to
the jack the main unit
or the power socket.
Incorrect installation
of nebulizer.
Incorrect installation of
nebulizer.
Pollution to the nebulizer
is serious.
The liquid can not be in
contact with nebulizing heat.
There are bubbles between
the liquid and nebulizing head.
Dirt electric pole of the
main unit.
Solutions
Replace all batteries.
Clean the nebulizer (please
refer to the section –
Cleaning and Disinfecting);
change nebulizing head
if necessary.
Reinstall batteries correctly.
Replace all batteries.
Reconnect it to power
socket or re-plug it into
the jack.
Reinstall the nebulizer
correctly.
Please refer to the section
– Assembling the Device.
Reinstall the nebulizer
correctly.
Clean the nebulizer; replace
nebulizing head if necessary.
Slightly tilt the main unit
towards yourself to make
the liquid gathers at the
nebulizing head side.
Gently shake the liquid.
Clean the electrode of the
main unit with medical gauze.
36