Section 10: Troubleshooting
Although service and repair of the Cavitron BOBCAT Pro
Ultrasonic Scaler should be performed by DENTSPLY
personnel, the following are some basic troubleshooting
procedures that will help avoid unnecessary service calls.
Generally, check all lines and connections to and from the
ultrasonic scaler. A loose plug or connection will often create
problems. Check the settings on the ultrasonic scaler's knobs.

10.1 Troubleshooting Guide

:
System will not operate
1. Check that the ON/OFF Control switch is in the ON ( I )
position, and that the detachable Power Cord is fully
seated in the receptacle on back of the ultrasonic scaler.
2. Check that the ultrasonic scaler's three-prong plug
is fully seated in an appropriate AC receptacle, and
that AC current is present.
Section 11: Warranty Period
The Cavitron Bobcat Pro Ultrasonic Scaler is warranted for ONE YEAR from date of purchase. Refer to the Warranty Statement
Sheet furnished with your system for full Warranty Statement and Terms.
Section 12: Specifications
Cavitron ® BOBCAT ® Pro: 115 VAC Systems
Electrical Voltage
115V Gen-130
Gen-130 (Model approved for U.S. application)
Current
0.8 amps @ 115V (80VA)
Frequency
50/60 Hertz
Water Pressure
25–60 psig
Flow Rate
Minimum setting (CW) = 0 ml/min
Maximum setting (CCW) > 60 ml/min
Weight
6 lb. (2.7 Kg)
Dimensions
Height: 8 in.
Width: 3.8 in.
Depth: 8 in.
Handpiece Cable: 96 in.
Foot Control Cable: 96 in.
Power Cord: 96 in.
Water Supply Line: 96 in.
ENGLISH • 10
System operates:
(No water flow to insert tip)
1. Assure that water control is properly adjusted.
2. Check that water supply control valve(s) (dental
office water supply) are open.
3. If connected to a DualSelect Dispensing System,
check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
(Insert stops vibrating)
1. Deactivate foot control.
2. Verify insert is 25kHz and in good condition.
3. Depress foot control.

10.2 Technical Support and Repairs

For technical support and assistance in the U.S. call
1-800-989-8826 or 717-767-8502 Monday through
Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time).
For other areas, contact your local DENTSPLY
representative.
Cavitron ® BOBCAT ® Pro: 220/230/240 VAC Systems
Electrical Voltage
220V Gen-130B
230V Gen-130U
240V Gen-130A, Gen-130E
Current
0.4 amps @ 220V/230V/240V (80VA)
Frequency
50/60 Hertz
Water Pressure
172-414 kPa
Flow Rate
Minimum setting (CW) = 0 ml/min
Maximum setting (CCW) > 60 ml/min
Weight
2.7 kg
Dimensions
Height: 20.3 cm
Width: 9.6 cm
Depth: 20.3 cm
Handpiece Cable: 2.4 m
Foot Control Cable: 2.4 m
Power Cord: 2.4 m
Water Supply Line: 2.4 m
OPERATING ENVIRONMENT
Temperature
10 to 25 degrees Celsius
Relative Humidity
30 to 75 percent ( non-condensing)
TRANSPORT AND STORAGE CONDITIONS
Temperature
0 to 70 degrees Celsius
Relative Humidity
10 to 95 percent ( non-condensing)
Atmospheric Pressure
500 HPa to 1060 HPa
SYMBOLOGY
AC POWER
TYPE B APPLIED PART EQUIPMENT
PROTECTIVE EARTH (GROUND)
Footswitch not for operating theatres
IPX1
Protection Class- IPX1
IPX1 Classification of ingress of water
Section 13: Equipment Classifications
• Type of protection against electric shock:
• Degree of protection against electric shock:
• Degree of protection against the harmful ingress of water:
• Mode of operation:
• Degree of safety of application in the presence of a flammable anaesthetic
mixture with air or with oxygen or nitrous oxide:
• According to medical device directive:
Section 14: Disposal of Unit
• In accordance with local and state laws.
Section 12: Specifications cont'd
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE WITH
UL-2601-1/60601-1, CAN/CSA C22.2 NO.601.1
13VA
Refer to instruction
manual for information
The International EMC-Mark
Proof of product compliance
with EMC directive (89/336/EEC)
Class 1
Type B
Ordinary
Continuous
Equipment not suitable for use in the presence
of flammable anaesthetic or oxygen.
IIA (rule 9)
ENGLISH • 11