11. Tighten the high-pressure
outlet/L-piece
screw
connection
(6)
using
a hexagon socket (wrench
size 14).
12. Tighten
the
L-piece/
200 bar
module
screw
connection (7) using a
hexagon socket (wrench
size 14).
13. Tighten
all
screw
connections
on
200 bar option.
14. Tighten the 2 screws on
the bracket.
NOTICE
i
i
Dräger
also
recommends
reducer R 57 584 to the high pressure input.
This makes it possible to ensure that the air is only fed
into the test device with a pressure of 200 bar.
Fitting the adapter for mask/helmet combinations
1. Use the drill supplied to drill a hole at each of the points
marked on the test head.
NOTICE
i
i
The "top" of the test head is approx. 5 mm thick,
the test head is hollow inside. The interior of the test
head cannot be damaged by the drilling.
2. Align the adapter and
fasten with the two screws.
3.3.3
Setting up the test device
NOTICE
i
i
Position the test device on a table or workbench.
The PC should be positioned next to the test device so
that it is within the operator's reach.
Dräger Quaestor 5000/7000
the
connecting
pressure
1. Align
the
parallel with the table edge.
The distance between the
table edge and the front
holes on the table mount
6
must be 12 cm.
7
2. Fasten the table mount to
the table using the screws
supplied.
3. Raise
the
slightly at the back.
01321868.eps
4. Align the front feet on the test device with the slots in the
table mount and move the test device backwards in the
table mount. The back feet must engage with the table
mount.
5. Ensure
that
screw (1)
on
pressure inlet is closed.
WARNING
!
Compressed air pipework may only be connected by
trained personnel.
Failure to follow this warning could lead to death or se-
rious injury.
CAUTION
!
For the test device, only breathing air according to
EN 12 021 from a stationary compressed air network
supply or from a compressed air cylinder may be used.
If the test device is inflated with oxygen, there is a risk
of fire.
The maximum permissible high pressure is 300 bar
(4500 psi). The maximum permissible medium
02321868.eps
pressure is 20 bar (58 psi). At higher pressures,
both the test device and the breathing safety product
may be damaged.
6. Connect the compressed air pipework or the compressed
air hose to the high pressure inlet.
7. Connect the test device to the PC using the USB cable.
8. Connect the test device to the power supply:
Align the red marking on the power pack plug and on the
test device with one another, and insert the power pack
plug into the test device.
Connect the power supply unit to the mains.
NOTICE
i
i
To disconnect the plug, move the socket approx. 1 mm
away from the test device. This releases the lock,
and the plug can be removed from the test device.
Commissioning the test device
table
mount
test
device
the
vent
the
high
12 cm
01521868.eps
1
01621868.eps
19