Specifications; Electrical - Leep System 1000 Workstation - Cooper Surgical LEEP System 1000 KH1000A Manual De Funcionamiento

Extractor de humos
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 69
Recommended separation distance between portable and mobile RF communications equipment
and the LEEP System 1000
The Electrosurgical Generator is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Electrosurgical Generator can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Electrosurgical Generator as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power
of transmitter
Watts
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output rating of the transmitter in watts (W) according to the
transmitter manufacturer.
For the Electrosurgical Generator,
V
= 3 Vrms
1
E
= 3 V/m
1
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Section 19 Specifications
19.1 Electrical – LEEP System 1000
INPUT VOLTAGE . . . . . . . . . . . . . . . . . . 190 –250 VAC, 50 Hz
CURRENT . . . . . . . . . . . . . . . . . . . . . . . 3.0 amps max.
FUSES . . . . . . . . . . . . . . . . . . . . . . . . . . 8.0 amps, 250 V, T type
WORKING FREQUENCY . . . . . . . . . . . . 450 kHz
OUTPUT POWER . . . . . . . . . . . . . . . . . 100 watts RMS (500 ohm load)
DUTY CYCLE . . . . . . . . . . . . . . . . . . . . . Intermittent 10/30 seconds
LOW FREQUENCY LEAKAGE . . . . . . . Less than 50 micro-amps
POWER CORD . . . . . . . . . . . . . . . . . . . . Various
WEIGHT . . . . . . . . . . . . . . . . . . . . . . . . . Approximately 98 pounds (44 kg)
®
LEEP System 1000
Workstation • Model KH1000A • Operating Manual • English
®
Electrosurgical Generator.
150 kHz to 80 MHz
0.1167
0.3689
1.1667
3.6894
11.667
®
Workstation
Separation distance according
to frequency of transmitter
Meters
80 MHz to 800 MHz
0.1167
0.3689
1.1667
3.6894
11.667
(continued)
800 MHz to 2.5 GHz
0.2333
0.7379
2.3333
7.3789
23.333
27

Publicidad

Tabla de contenido
loading

Productos relacionados para Cooper Surgical LEEP System 1000 KH1000A

Tabla de contenido