19.2 General Specifications – LEEP System 1000
ENVIRONMENTAL CONDITIONS
(Usage)
ENVIRONMENTAL TEMPERATURE . . . Between 50 °F and 113 °F (10 °C and 45 °C)
RELATIVE HUMIDITY . . . . . . . . . . . . . . Between 30% and 75%
(Shipping and Storage)
ENVIRONMENTAL TEMPERATURE . . . Between -40 °F and 158 °F (-40 °C and 70 °C)
RELATIVE HUMIDITY . . . . . . . . . . . . . . Between 10% and 100%
PHYSICAL - Electrosurgical Generator
DIMENSIONS . . . . . . . . . . . . . . . . . . . . 13" wide x 14" deep x 7¼" high (330 mm x 356 mm x 184 mm)
WEIGHT . . . . . . . . . . . . . . . . . . . . . . . . . 30 pounds (14 kg)
ELECTROSURGICAL OUTPUT
RF OUTPUT FREQUENCY . . . . . . . . . . 450 kHz
RF OUTPUT POWER:
CUT
0-100 watts RMS*
BLEND
0-100 watts RMS*
COAG
0-80 watts RMS*
*Stable to >800 ohms (calibration at 500 ohms)
RF ISOLATION . . . . . . . . . . . . . . . . . . . . Less than 150 milliamps at 200 ohms
CLASSIFICATION . . . . . . . . . . . . . . . . . I-Type BF
OUTPUT CIRCUIT . . . . . . . . . . . . . . . . . Floating output. Protected against the effects of the defibrillator.
WORKING MODE . . . . . . . . . . . . . . . . . Discontinuous maximum duty cycle: 10/30 sec.
COOLING . . . . . . . . . . . . . . . . . . . . . . . . Convection cooling without fan
CONTROL . . . . . . . . . . . . . . . . . . . . . . . Foot pedal operated (pneumatic) with audible signals
AUDIBLE SIGNALS AND LIGHTS FOR OPERATION AND ALARM:
MAIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . Green light
ALARM, PATIENT PLATE CONTINUITY . Low pitch intermittent audible alarm - red light
ALARM, OUTPUT POWER . . . . . . . . . . Higher pitch intermittent audible alarm
CUT AND BLEND MODES . . . . . . . . . . . Low pitch audible signal - yellow light
COAGULATION MODE . . . . . . . . . . . . . High pitch audible signal - blue light
NOTE: Specifications subject to change.
OUTPUT POWER AT 500 OHMS:
Pure Cut
100 W RMS (open circuit 830 Vp-p; crest factor 1.43)
Waveform: sinusoidal at 450 kHz
Blend Cut
100 W RMS (open circuit 1200 Vp-p; crest factor 2)
Waveform: sinusoidal at 450 kHz
Duty cycle: 60%
Coagulation
80 W RMS (open circuit 3800 Vp-p; crest factor 5.5)
Waveform: sinusoidal at 450 kHz
Duty cycle: 10%
28
Volts p-p Max
(open circuit)
830
1200
3800
and mode lights
®
LEEP System 1000
Workstation • Model KH1000A • Operating Manual • English
®
Electrosurgical Generator
Duty
Crest
Cycle
Factor
---
1.4
60%
2.0
10%
5.5
6
5
7
(continued)