28. Connect interface
Make sure the Unit is OFF before you attempt to connect the interface cable.
Use a cable with SUB-D 9 connectors and connect with the terminal connector located
underneath the LABOROTA 20 control control panel.
29. Interface parameters
9600 Baud, no parity, 8 bits, 1 stop bit, no protocol
30. RS 232 interface
This interface allows you to specify all Set-values, move the lift up and down, open and close
aerating valve and vacuum valve.
All Act-values and status of safety functions may be recalled.
31. RS 232 interface commands
query
Command
Act commands
VAC?
VAP?
TAU?
BATH?
CHILL?
SPEED?
TIME?
COUNT?
SET1?
SET2?
Safety functions
SC?
FS?
RS?
OP?
INTERFACE
action
remarks
Act Vacuum
Act Vapor
Act T auto
Act Bath
Act Chiller
Act Speed
Act Time
elapsed time, continuous operation
Act Time
time, timer mode
query Set-value
query Act-value
Safety Cover
0 = action; 1 = no action
Flask Support
0 = action; 1 = no action
Rotation Stop
0 = action; 1 = no action
Over
0 = action; 1 = no action
Temperature
Prot.
101