WARNING
Risk of deviating measured values during flow
measurement
Applies to Cato, Cicero, and Narkomed 6000:
Manual ventilation may result in deviating measured
values.

Replacing/Snapping into place

The disposable CO
absorber must be replaced as soon as one of
2
the following conditions applies:
The inspiratory CO
concentration has reached approx.
2
5 mmHg.
The soda lime in the absorber has turned 2/3 violet.
The disposable CO
absorber has been used for 4 weeks or
2
longer
NOTE
In contrast to Drägersorb 800 Plus, the color change of the lime
will persist while it is not in use and there is no regeneration of the
soda lime.
Always remove spent disposable CO
and dispose of immediately.
Replace the disposable CO
2
continuous CO
absorption in the breathing system.
2
1
Press the button: the disposable CO
2
Remove the disposable CO
3
Before fitting into place, shake the new disposable CO
absorber by turning it upside down several times in order to
loosen up the soda lime.
 Make sure to fit the disposable CO
correctly (pictograms must face the user).
4
Remove the red cap.
5
Slide the disposable CO
absorber onto the mounting as far as
2
possible.
10
absorbers from the device
2
absorber quickly to ensure
absorber swings out.
2
absorber from the mounting.
2
2
absorber into place
2
Instructions for use CLIC Absorber Free / Infinity ID CLIC Absorber Free / CLIC Adapter
6
Swing into the system until it snaps into place.
 The disposable CO
absorber must be fully covered by the
2
CLIC adapter.
CO
bypass function
2
WARNING
Risk of patient injury
There is no absorption of CO
disposable CO
absorber is swung out.
2
Activating the CO
bypass function
2
 Press the button: the disposable CO
Deactivating the CO
bypass function
2
 Swing the disposable CO
snaps into place.

Infinity ID function

Infinity ID CLIC Absorber with integrated transponder. For Dräger
devices using the Infinity ID technology, the transponder is used
for the device-side detection of the CLIC Absorber.
in the breathing system if the
2
absorber swings out.
2
absorber into the system until it
2