ty or have received instructions from them on how to
use the unit.
• The device must not be used in combination with elec-
tronic life-sustaining devices such as ventilators or
other electronic medical products, such as electro-
cardiographs.
• If your pain gets worse, becomes chronic, or persists
for up to five days, contact your doctor.
• Do not share this device with anyone else. This could
lead to skin irritation or infection. The device is intended
for private use only.
• Do not move or use the cuff if the control unit is
switched on.
• Only connect the connection cable for the electrode
plug to the intended control unit and the universal cuff
that is included.
• Do not pull on the connection cables during treatment.
• Do not bend or pull on the ends of the connection cable.
• Do not wear any electronic devices, such as watches,
while using the device.
• During treatment, do not touch the water contact elec-
trodes on the universal cuff with your finger.
• Do not use the device on a particular area more than
three times a day with a 30-minute treatment period
each time.
• If the device does not work properly, or you feel unwell
or experience pain, stop using it immediately.
• Do not modify the electrodes as this is potentially haz-
ardous (max. recommended output value for the elec-
trodes is 58.1 mA/sq in (9 mA/cm²), an effective current
intensity 12.9 mA/sq in (2 mA/cm²) requires increased
awareness).
• Do not use while undertaking any activity where an un-
expected reaction (e.g. strong muscle contractions even
at low intensity) could be dangerous.
• Do not use during strenuous activity (e.g. sports).
• Pre-treating the skin with greasy creams or lotions is
not recommended.
• At first, use the device while sitting or lying down to
minimize the risk of accidental injuries as a result of
feeling faint. If you feel faint, switch off the device im-
mediately, lie down and support the legs in an elevated
position (approx. 5 - 10 min).
• Hold the device away from sources of heat and do not
use it within three feet (approx. 1 m) of shortwave or
microwave devices (e.g. mobile phones).
• The device is intended for self-treatment only.
• If there are signs of wear and tear or damage or if the
device is used improperly, return it to the manufactur-
er or retailer.