MLED III Supercharged U7
The user must return the handpiece to its charging base
to recharge the unit's battery (see chapter 4.3
INSTALLATION).
VIII – MAINTENANCE
Before conducting any maintenance on the Mini
LED SuperCharged, check that:
- The Mini LED SuperCharged is not on its charging
base.
- The battery has been unscrewed from the
handpiece.
- The charging base has been disconnected from the
AC supply.
Before cleaning the handpiece, insert the protective
cap in the place of the light guide, supplied to ensure
that liquid does not enter the handpiece.
Avoid using cleaning and disinfection products
containing flammable agents (or other corrosive
agents such as acetone, chlorine or bleach).
Otherwise, ensure that the product completely
evaporates and that there are no combustibles on the
unit and its accessories before operation.
Do not use abrasive products to clean the unit.
Do not immerse the unit.
Never use ultrasound cleaning, whether for the Mini
LED SuperCharged or its accessories.
None of the accessories are delivered in a sterilized
state.
Only the light guide and the flexible protective shield
are sterilizable.
Before sterilization, check the cleanliness of your
autoclave and the quality of the water used.
After each sterilization cycle, immediately remove
the items from the autoclave to reduce the risk of
corrosion of metallic parts.
It is necessary to leave the sterilized items to cool
down to the ambient temperature and dry before re-
using them.
It is recommended to sterilize the items identified as
sterilizable by Satelec individually in the sterilization
bags foreseen for this purpose.
To maintain the sterile and aseptic state of
accessories, make sure that they are kept in
hermetically-sealed bags or containers suited to use
V6:Mise en page 1
05/09/2011
in dentistry.
The maintenance and/or sterilization instructions
that follow must be applied before each use of the
unit.
8.1 – Pre-disinfection/cleaning
Clean and disinfect the body, light guide, flexible shield
and rigid protection shield, and the charging base of the
Mini
cleaning/disinfecting wipes based on alcohol,
amphoteric disinfectant and biguanide (refer to the
manufacturer's instructions) for at least two minutes.
Leave the product to act for at least 15 minutes.
Use wipes with CE marking, or compliant with any
standard that may be required by national regulations.
8.2 - Drying
Dry using a clean, single-use, non-woven cloth to
remove any liquid traces.
8.3 - Packing
Pack in single-use sterilization bags or sleeves compliant
with the requirements defined in the ISO 11607-1
standard, or compliant with any standard that may be
required by national regulations.
8.4 - Sterilization
The optic fiber and the flexible protective shield for the
Mini LED SuperCharged must be sterilized individually in
an autoclave according to the following parameters:
- Autoclave: Type B compliant with the EN 13060
- Sterilization temperature: 134°C.
- Sterilization steady state: 18 minutes.
- Pressure: 2 bar minimum.
CAUTION: Not all autoclaves can reach 134°C. Not all
autoclaves perform a pre-depressurization. For further
information on the applicable sterilization instructions,
consult the autoclave manufacturer.
8.5 - Storage
After this, store the sterilized items in a dry, dust-free
place.
Before re-use, in the event of nonconforming packaging
integrity, re-package and re-sterilize according to the defined
protocol.
12
09:53
Page 12
LED
SuperCharged
standard.
using
ready-to-use