IX - TROUBLESHOOTING
In the event of a problem, before contacting the after-sales service of SATELEC or the supplier:
- Ensure that the base is correctly connected to the AC supply to ensure that the battery is charged normally. If the
Mini LED SuperCharged s holder is not lit despite being correctly connected, contact the after-sales service of
SATELEC or the supplier.
- Check that at least one of the four round battery charge level indicators is visible on the LCD screen before pressing
the ON/OFF button.
- In the event of a faulty battery (see VII – INTERLOCKS) an automatic interlock system will prevent the unit from
operating. If this is the case, recharge the battery by placing the lamp on its holder or by using a second fully-
charged battery.
- Intensive use of the lamp can result in high temperatures inside the unit. If this happens, an automatic interlock
system prevents the unit from operating (see VII – INTERLOCKS). Leave the unit at rest for a few minutes to allow
it to cool down.
- After each use, check that there is no composite residue adhering to the light guide. If this is the case, immediately
remove the residues and check that the surface of the light guide has not been damaged. If damage is visible,
replace the light guide as the lamp's power could be significantly reduced.
- Under normal conditions of use, the lamp's power does not vary if the battery is properly charged. Consequently, it
is not necessary to check power as with ordinary polymerization lamps. Nevertheless, in case of doubt, check its
power using your light guide.
- The battery located at the bottom of the handpiece can be replaced by unscrewing it.
- After removing the light guide, check that the LED is clean and undamaged. If necessary, clean it using a medical-
quality dry air jet (free from compressor oil).
- During verification of the power level of the handpiece, if the translucent handpiece holder base is red, check if
the LED, light guide and verification window are clean.
- In the event of dust being present, clean using a dry air jet.
- If the problem persists or if the light guide or the verification window are damaged, the lamp must be returned to
the after-sales service.
In the event of an anomaly, contact the supplier of the unit rather than just any repairer, which may return
your unit in a hazardous state for both you and your patients.
The technical service of your supplier is at your disposal for any technical problems encountered on your unit.