Use
2.6
Marking
The following information is shown on the product label:
Date of manufacture
Serial number
Order number
Approval marking
Testing standard
Name of device
Nominal usage period of the device (see section 2.2 on page 7)
3
Use
3.1
Before initial use
Before commissioning, the delivery date must be inscribed on the product label.
The use of punch numbers for the inscription is prohibited. Dräger recommends
engraving or etching.
3.2
Using the oxygen self-rescuer
WARNING
!
Keep the oxygen self-rescuer away from combustible substances
(petrol, grease, solvents, etc.) before or during use to prevent a
potential risk of fire!
Destroying the oxygen self-rescuer can result in a potential risk of
ignition if the oxygen-abstracting chemical comes into contact with
combustible substances such as charcoal.
Regularly practice using the oxygen self-rescuer with a training device
The oxygen self-rescuer might be difficult to open in cold weather
conditions, and especially at minus temperatures.
Only open the oxygen self-rescuer when using it!
Otherwise moisture from the ambient air will flow into the open oxygen self-
rescuer and impair its function.
Opened devices cannot be closed again. They are considered used and
must be disposed of, not stored, see section 7 on page 10.
Check the oxygen self-rescuer prior to carrying/daily see section 3.3 on
page 8.
Use the oxygen self-rescuer only once.
3.3

Preparation for use

The following work steps must be performed every day. If one of the criteria listed
is not met, the oxygen self-rescuer must not be used.
NOTICE
i
i
If the oxygen self-rescuer suffered unusual strain (blows, pressure,
etc.), the checks mentioned should be carried out immediately.
In the following cases, the oxygen self-rescuer may be repaired by trained
personnel:
If the oxygen self-rescuer is worn with a shoulder harness or hand strap,
check that they are locked onto the support ring. If the carrying strap is
frayed or damaged, it must be replaced.
Make sure that the viewing window is not damaged. A damaged viewing
window must be replaced.
In the following cases, the oxygen self-rescuer must be taken out of service or
sent to Dräger for inspection.
Check that the oxygen self-rescuer is complete.
Check that the sealing tag is not damaged.
Check that the housing is closed.
Check that the opening mechanism is locked and undamaged.
Check the housing for deformations (see supplementary sheet).
Ensure that the housing does not have any tears or holes.
Check the viewing window: If numerous yellow fragments, the same size or
larger than grains of sugar, can be shaken in the viewing window area, the
oxygen self-rescuer must be decommissioned.
An adhesive, thin yellow dust layer is normal after years of carrying and
does not affect the function.
1
not part of the EC type examination
8
Check if the moisture indicator
(1) is brown. If moisture has
penetrated into the oxygen self-
rescuer, the moisture indicator
changes
from
turquoise.
If the oxygen self-rescuer is fitted
with the optional heat indicator:
Check if the heat indicator (2) is
light-coloured.
The
indicator's colour will turn dark if
the device has been exposed to
too much heat.
3.4
Donning procedure
NOTICE
i
i
Donning the device incorrectly may cause a delay when using the
oxygen self-rescuer in emergencies.
Be sure to perform the following steps in the described sequence.
1.
Pull the opener up and press it
into
its
end
position.
tightening straps then release
automatically.
2.
Remove the upper part of the
housing and dispose of it.
Opening can be supported as
1
.
follows:
Use the fully opened opener
to lever the lid open.
Or
Bend the housing at the
sealing surface. To do this,
place the device flat on the
floor. Push onto the edge of
the upper and
lower parts of the housing.
In
minus
temperatures,
support the opening with
your
feet
if
(analogous to the previous
point).
3.
Grasp the yellow loop and use it
to pull the oxygen self-rescuer
out of the housing.
WARNING
!
Do not pull the oxygen self-rescuer at its mouthpiece, at the hose or at
other parts out of its housing, but only at the yellow loop. Observe the
indicated direction when pulling it out.
Otherwise, the oxygen self-rescuer could be damaged and may not
supply enough breathing air.
1
brown
to
heat
2
The
necessary
Dräger Oxy 3000/6000 MK III
00931209.eps
01031209.eps
00300133.eps