ATTENTION! WE DO NOT ADMIT EQUIPMENTS TO REPAIR IF THEY ARE
NOT WELL DISINFECTED.
9. TROUBLE SHOOTING
PROBLEM
CAUSE
Digits are irregular
Wrong connection of the
electrode
Irregularity in the measu-
Dirty or damaged electro-
rements
de
Display
shows
"Err"
Anomaly during check
when on
Display shows "Err" in
Input potential exceeds
mV value and right tempe-
measurement range
rature value
Button "pH/mV" does
When measurement pH
not let select measurement
range exceed ±20.00 dis-
pH mode
play indicates "Err" and
continues to mV measure-
ment mode
Version 3 August 09
SOLUTION
Check the connections
Clean electrode
Replace electrode
Turn off the equipment,
wait for a while and turn on
again. If the problem per-
sists, please, contact the
Technical Service
Insert protector device in
the connection of the elec-
trode to re-establish the
system
Reduce input signal
Insert protector device to
the electrode connection to
re-establish the system
Instruction manual 50903000
ENGLISH
INDEX OF LANGUAGES
Spanish
English
French
INDEX OF CONTENTS
1.FEATURES OF THE EQUIPMENT
2.THEORETICAL FUNDAMENTALS
3.DESCRIPTION
4.TECHNICAL SPECIFICATIONS
5.INSTALLATION / SETTING UP
6.CALIBRATION
7.Ph-METER FUNCTIONS
8.MAINTENANCE AND CLEANING
9.TROUBLE SHOOTING
10.ACCESSORIES
APPENDIX I: Diaphragm types of referent electrode / Features
CE CERTIFICATE
1. FEATURES OF THE EQUIPMENT
Digital pH-meter is indicated to make measurings of pH at compensating temperature
and electrochemical potential (mV) of watery solutions, so we can get diverse information,
for example:
Direct measurement of concentrations of anions and cations.
Direct measurement of dissolved gases as NH 3 , O 2 , CO 2
Acidity and alkalinity in water
Calcium in water
pH in wter, slurries, suspensions
Proteins in water
Its features extend to many fields of work, we emphasized the following:
Page 29
Page 18
Instruction manual 50903000
ENGLISH
2-16
17-30
31-45
18
19
20
21
21
23
26
27
29
30
30
46
Version 3 August-09