4. Use only Nellcor wraps designed for use with this sensor. Do not use tape. Use of additional
tape or various other types of wraps can cause skin damage.
5. Intravascular dyes or externally applied coloring such as nail polish, dye, or pigmented cream
may lead to inaccurate measurements.
6. Excessive motion may compromise performance. In such cases, try to keep the patient still,
or change the sensor site to one with less motion.
7 When the sensor is wrapped too tightly or supplemental tape is applied, venous pulsations
may lead to inaccurate saturation measurements.
8. As with all medical equipment, carefully route cables to reduce the possibility of patient
entanglement or strangulation.
9. Using excessive force when removing the disposable wrap may damage the sensor.
10. Do not sterilize by irradiation, steam, or ethylene oxide. Such sterilization could damage the
sensor.
11. Do not alter or modify the SpO
or accuracy.
12. For additional warnings, cautions or contraindications when using this sensor with
Nellcor-compatible instruments, refer to the instrument operator's manual or contact the
manufacturer of the instrument.
Note: High oxygen levels may predispose a premature infant to develop retinopathy. Therefore,
the upper alarm limit for oxygen saturation must be carefully selected in accord with
accepted clinical standards and considering the accuracy range of the oximeter being
used.
If you have questions regarding any of this information, contact Covidien Technical Services
Department, or your local Covidien representative.
Warranty
To obtain information about a warranty, if any, for this product, contact Covidien Technical
Services or your local Covidien representative.

Accuracy Specifications

For the accuracy specification range when used with Nellcor™ monitors, refer to information
provided with the monitor, or (in the U.S.), contact Covidien Technical Services Department.
Outside the U.S., contact your local Covidien representative.
For the accuracy specification range of this sensor when used with Nellcor-compatible
instruments, refer to the instrument operator's manual or contact the instrument manufacturer.
Use this sensor only with Nellcor instruments and instruments containing Nellcor oximetry,
or with instruments licensed to use Nellcor sensors (Nellcor-compatible instruments).
This sensor integrates Nellcor OxiMax™ technology into its design. When connected to
an OxiMax-enabled instrument, this sensor uses OxiMax technology to provide additional
Sensor. Alterations or modifications may affect performance