Calibration Procedure
NOTE: This procedure should be performed in the same
geographic area as the measurements for infant
weights. Procedure requires two certified 10 kg
Class F test weights.
1. Press
button while the scale is off
1
2. Press
button while the display
is flashing all segments --> Display
shows "Func 1"
reweigh
3. Press
fed baby
button --> Display
shows "Func 2"
4. Press
button --> Display
3
shows "9XXX" (For example, 9802
will be displayed)
NOTE: The gravity setting will update automatically upon
completing the calibration procedure.