Calibration Procedure (cont.)
5. Press
button again -->
3
Display shows "20 0--" and "g"
Wait for display to show "20 2--"
6. Place two 10 kg weights on the
platform after the display shows
"20 2--"
7. Wait for the display to show
"stabilized" mark (circle in the top
of left corner), then Press the
button to complete calibration
8. Wait for display to show "func 1"
then remove the 20 kg weight
off
9. Press
button to turn off the scale
NOTE: If the following error messages are displayed, "Er0065"
or "Er0066", recalibration must be performed.
CAUTION: Do not leave the calibration weights on the scale
for prolonged periods of time (more than a few
minutes)