Fluids; Correct Use; Use; Range Of Use (Indoor Use Only) - IKA IC basic Instrucciones De Manejo

Ocultar thumbs Ver también para IC basic:
Tabla de contenido

Publicidad

• After a power failure during operation, the device may
start automatically (depending on operating mode).
• Transport the device with care.
• Do not transport or empty the bath while it is still hot.
This may result in accidents, especially scalding injuries.

Fluids

Only use fluids, which fulfill the re-
DANGER
quirements for safety, health and
device compatibility. Be aware of
the chemical hazards that may be associated with the
bath fluid used. Observe all safety warnings for the fluids.
• Depending on the bath fluid used and the type of opera-
tion, toxic or flammable vapors can arise. Ensure suit-
able extraction.
• Do not use any fluid which may cause dangerous reac-
tions during processing.
• Only use recommended bath fluid. Only use non-acid
and non corroding fluid.
Only process and heat up any fluid
DANGER
that has a flash point higher than
the adjusted safe temperature limit
that has been set. The safe temperature limit must al-
ways be set to at least 25 ºC lower than the flash point
of the fluid used. Examine regularly the function of the
safety temperature limiter.
Never operate the device without
CAUTION
sufficient heat carrier fluid! You
should also be careful to ensure
that the minimum clearances and immersion depths in
the fluid are observed. Check the fluid level detection at
a regular basis (see the chapter "Filling and draining").
• Continuous monitoring of the bath and the filling level of
the bath fluid is required, especially at high temperatures.
• For optimum temperature stability, the fluids viscosity
should be 50 mm
2
/s or less at its lowest operating tem-
perature. This permits good fluid circulation and mini-
mizes heating from the pump.

• Use

Use IC (Immersion Circulator) and HBC (Heating Bath
Circulator)
- For heating and circulating fluids
Intended Use: Tabletop Device

• Range of use (indoor use only)

- Laboratories
- Schools
- Pharmacies
- Universities
CAUTION
tion steam.
• Untreated tap water is not recommended. It is recom-
mended to use distilled water or high purity water (ion
exchangers) and add 0,1 g soda (sodium carbonate
Na
CO
) /liter, to reduce corrosive properties.
2
3
DANGER
WARNING
CAUTION
WARNING
- Untreated tap water
- Concentration of acids or bases
- Solutions with halides: chlorides, fluorides, bromides,
iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.
DANGER
move the remaining water from the complete system
(including hoses and external devices). When doing this,
also open the stopper and union nuts caps of the pump
outputs and inputs and blow compressed air through
all the pump outputs and inputs! Beware of the risk of
burning due to delay in boiling!

Correct use

This device is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply
network that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not
supplied or recommended by the IKA
- If the device is operated improperly or in contrary to the
IKA
specifications.
®
- If the device or the printed circuit board are modified by
third parties.
If water is used at higher tempera-
ture, there is heavy loss of fluid due
to the condensation of evapora-
Risk of burning caused by vapor or
hot water at the outlet of the cool-
ing coil.
Do not use the cooling coil with
water at bath temperatures > 95°
C.
For bath temperatures > 60°C make
sure that the flow rate through the
cooling coil is high enough.
Don't use following fluids:
When changing the bath fluid
from water to a heat transfer fluid
for temperatures above 100 °C, re-
®
.
25

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hbc 5 basicHbc 10 basic

Tabla de contenido