The following processes are started automatically: 
Water intake: 
The tank takes in water through the nozzles (approx. 6 litres) 
Heating: 
The cleaner heats the water and performs an initial decontamination of 
the instruments for 5 minutes. 
If the preset temperature is not reached, heating will continue during 
cleaning. 
Disinfectant/detergent intake % and mixing the solution:
The pump inside the cleaner lets in 0.12 litres of disinfectant/detergent. 
 Disinfectant loading / manually detergent: * 
Select "No" in the program "Disinfectant entrance/ cleaner. Choose the 
other parameters and press  
water loading ends the device beeps: open the door and put the 
disinfectant in the tank. Opening  the door puts in pause the cycle until 
the closing. 
In case of non‐opening door within 60 seconds, the device will continue 
the cycle with the next step. 
*  Procedure to be followed when you are using a liquid different from the 
one entered in the tank 
Degassing:
This process removes the gases dissolved in the water. 
BLINKING
Ultrasonic cleaning:
Cleaning by cavitation. 
Discharge:
The water is discharged. 
Rinse:
The cleaner rinses the instruments for few minutes to remove any trace 
of disinfectant/detergent.
 
 
 
 
  to start the cycle. Once the process of 
 
17
ENGLISH
 