Setting The Safe Temperature Limit - IKA RO 5 Instrucciones De Manejo

Ocultar thumbs Ver también para RO 5:
Tabla de contenido

Publicidad

Unpacking
• Unpacking
- Please unpack the device carefully
- In the case of any damage a detailed report
must be sent immediately (post, rail or forwarder)
Correct use
• Use
- For mixing and / or heating liquids
• Range of use
Indoor environments similar to that a laboratory of research,
teaching, trade or industry area.
Commissioning
RO/RT
Observe the ambient conditions (temperature, humidity, etc.)
listed under Technical Data.
The device is switched on and off via the switch (A) located on
the right-hand side of the device. On the front panel a small
arrow labelled Power above the display (B) indicates the location
of the switch.
Safe temperature limit
RT
The max. attainable heating plate temperature is restricted to 130
°C by a permanently set safety temperature limit. Once this limit
has been attained, the device switches off the heating.

Setting the safe temperature limit

RT
The safety temperature is set as follows:
Switch the device on using the On/Off switch
on the right side of the device.
As soon as the display shows SAFE, press and
hold down the Temp membrane key and use
Stirring function
RO/RT
The stirring function is started and stopped by pressing the „Start/
Stop" button (C).
Pressing the +/- buttons (E or D) increases or reduces the speed.
• Delivery scope
RO
RT
- Magnetic stirrer
- Heating magnetic-
RO 5/10 or 15
stirrer RT 5/10 or 15
- power supply unit
- mains cable
- operating instruction
- operating instruction
The safety of the user cannot be guaranteed:
- if the device is operated with accessories that are not supplied
or recommended by the manufacturer,
- if the device is operated improperly or contrary to the
manufacture's specifications,
- if the device or the printed circuit board are modified by third
parties.
After the device is switched on a display self-test is performed.
The display cycles through the following information:
• All segments light up
• The software version and the selected mode are displayed
• The selected "Eco Mode" is displayed
(E when ECO mode is active)
• If the drive has not been started the display reads „OFF". If
the drive has been started the selected speed is displayed.
The safe temperature limit must all-
WARNING
ways be set at least 25 °C lower than
the fire point of the media to be
processed!
the +/- membrane keys (H or I) to set the
desired safety temperature.
The set safety temperature is displayed.
Once the Temp membrane key is released, the safety temperature
is saved.
11

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ro 10Ro 15Rt 5Rt 10Rt 15

Tabla de contenido