Problem
After switching on for
the first time, measure-
ment results are trans-
mitted and two beeps
are heard.
Only the "SYS" item is
visible in the rf menu.
Only the "SYS" and "lrn"
items are visible in the rf
menu.
The items "APrt" and
"Time" are not visible in
the rf menu.
After calling up the
menu, the "rf" item is not
displayed.
The display "Er:X:11"
appears.
The display "Er:X:12"
appears.
90 •
Cause / Solution
• The device was unable to send measurement
results to the wireless receiver (seca wireless
printer or PC with seca USB wireless adapter).
– Ensure that the scale is integrated in the
wireless network.
– Ensure that the receiver is switched on.
• Reception is being interfered with by HF equip-
ment (e.g. cell phones) in the vicinity.
- Keep HF equipment at least 1 meter away
from transmitters and receivers in the seca
wireless network. The actual transmission
output of HF equipment may require mini-
mum distances of more than 1 meter. For
details, go to www.seca.com.
NOTE
If this fault is not eliminated, there will be no
fresh acoustic warning when further
attempts are made to send data.
• The wireless module is deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 84).
• The wireless module is activated and no wire-
less group is set up.
- Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 84).
• No wireless printer registered to the wireless
group.
- Register the wireless printer to the wireless
group via the "lrn" menu item (see "Setting up
a wireless group (Lrn)" on page 84).
• The scale's wireless module is defective.
- Inform seca service.
The scale has too high a load or too high a load in
one corner.
- Take the load off the scale or distribute the
weight more evenly.
- Restart the scale.
The scale has been switched on with too high a
load.
- Remove the load from the scale.
- Restart the scale.