8.2.
Voltagemeasuring
-
U sing the rotary switch select the required function (V AC) or (V DC) on
the BENNING MM 8/ 9/ 10.
-
T he black safety measuring cable has to be connected to the COM jack
on the BENNING MM 8/ 9/ 10.
-
The red safety measuring cable has to be connected to the jack for V, Ω,
µA,
, Hz on the BENNING MM 8/ 9/ 10.
-
B ring the safety measuring lines into contact with the measuring points, read
off measured value on the digital display on the BENNING MM 8/ 9/ 10.
see figure 2:
see figure 3:
8.2.2 Currentmeasuring
-
U sing the rotary switch select the required range and function (µA DC
B ENNING MM 8/ 9/ 10, A AC or A DC BENNING MM 9/ 10).
-
T he black safety measuring cable has to be connected to the COM jack
on the BENNING MM 8/ 9/ 10.
-
The red safety measuring cable has to be connected to the jack for V, Ω,
µA,
, Hz (direct currents on the 6 mA) on the BENNING MM 8/ 9/ 10
or to the jack for the 10 A range (direct or alternating currents up to 10 A)
on the BENNING MM 9/ 10.
-
B ring the safety measuring lines into contact with the measuring points, read
off measured value on the digital display on the BENNING MM 8/ 9/ 10.
see figure 4:
see figure 5:
8.3 Resistancemeasurements
-
U sing the rotary switch select the required function (Ω) on the
B ENNING MM 8/ 9/ 10.
-
T he black safety measuring cable has to be connected to the COM jack
on the BENNING MM 8/ 9/ 10.
-
The red safety measuring cable has to be connected to the jack for V, Ω,
µA,
, Hz on the BENNING MM 8/ 9/ 10.
-
B ring the safety measuring lines into contact with the measuring points, read
off measured value on the digital display on the BENNING MM 8/ 9/ 10.
see figure 6:
8.4 Diodetest
-
U sing the rotary switch select the required function (
BENNING MM 8/ 9/ 10.
-
T he black safety measuring cable has to be connected to the COM jack
on the BENNING MM 8/ 9/ 10.
-
The red safety measuring cable has to be connected to the jack for V, Ω,
µA,
, Hz on the BENNING MM 8/ 9/ 10.
-
B ring the safety measuring lines into contact with the diode con-
nections, read off measured value on the digital display on the
B ENNING MM 8/ 9/ 10.
-
F or a standard Si diode applied in conduction direction the conduction volt-
age between 0.400 V to 0.900 V is displayed. The display "000" indicates a
short circuit in the diode, display values "OL" indicate an interruption in the
diode.
-
F or a diode applied in reverse direction "OL" is displayed. If the diode is
defective, "000" or other values are shown.
see figure 7:
8.5 Continuitytestingwithbuzzer
-
U sing the rotary switch select the required function (
B ENNING MM 8/ 9/ 10.
-
T he black safety measuring cable has to be connected to the COM jack
on the BENNING MM 8/ 9/ 10.
-
The red safety measuring cable has to be connected to the jack for V, Ω,
µA,
, Hz on the BENNING MM 8/ 9/ 10.
-
T he safety measuring lines have to be brought into contact with the meas-
uring points. If the line resistance between the COM jack and the jack
for V, Ω, µA,
B ENNING MM 8/ 9/ 10.
see figure 8:
11/ 2006
Direct voltage measurement
Alternating voltage measurement
Direct current measurement (µA range, BENNING MM 8)
Alternating current measurement (BENNING MM 9/ 10)
Resistance measurements
Diode test
, Hz is below 100 Ω, the fitted buzzer sounds on the
Continuity testing with buzzer
B ENNINGMM8/9/10
) on the
) on the