User Controls
ON / OFF Button
Press once to turn "ON"; Press and hold for one second to turn "OFF".
Audible Alarm Button
Pressing this button will toggle the Inogen One® G3's breath detection
audible alert on and off:
1. Default Mode. When the Inogen One® G3 is powered up, the breath
detection audible alert is disabled. The display's mode indication area
will show a bell icon crossed out with an X in default mode.
2. Breath Detection Alert Mode. The Inogen One® G3 will alert with
audible and visual signals for "no breath detected" when this mode
is enabled and no breath has been detected for 60 seconds.
At 60 seconds, the device will enter into auto pulse mode and once
another breath is detected, the device will exit auto pulse mode and
deliver normally on inspiration. The display's mode indication area will
show a bell icon, flashing red light and display message when the
alert is enabled.
3. If power is lost, breath detection alert mode resets to
default mode.
Flow Setting Control Buttons
Use the – or + flow setting control buttons to select the setting
as shown on the display. There are four settings, from 1 to 4.
Display Backlight Button
Push on; automatically turns off after 10 seconds.
User Interfaces
Display
This screen displays information regarding flow setting, power
status, battery life and errors. If you would like to change the
language on the Inogen LCD Screen contact Inogen Client Services.