Aesculap Suhl GmbH
Clipping machine Favorita Hybrid/FAV5 Hybrid
5.
Processing
Damage to, or destruction of the
product
cleaning/disinfecting or steriliza-
tion!
CAUTION
Only clean and disinfect the
product manually.
Do not sterilize the product
under any circumstances.
5.1 Cleaning/disinfection
Note
Always follow disposal and hygiene instructions and
regulations!
Manual cleaning/disinfection
Risk of electric shock and fire haz-
ard!
Unplug the device before clean-
DANGER
ing.
Ensure that no fluids will pene-
trate the product.
Leave the product exposed to
air for at least 1 minute after
cleaning/disinfecting.
Damage to the product due to
incorrect cleaning or inappropriate
cleaning/disinfecting agents!
CAUTION
Use any commercial detergent
for surface cleaning.
Never immerse the product in
water or detergent.
8
caused
by
mechanical
Wipe the shell of the product with a lint-free cloth
moistened with a commercial disinfectant for wipe
disinfection.
If necessary, after the appropriate exposure time,
wipe off any residues of the cleaning/disinfecting
agent with a lint-free cloth moistened with clean
water.
Dry the product with a clean, lint-free cloth.
If necessary, repeat the cleaning/disinfecting pro-
cess.
5.2 Control, care and inspection
Allow the product to cool down to room tempera-
ture.
Inspect the product after each cleaning and disin-
fecting cycle to be sure it is: clean, functional, and
undamaged.
Lubricate the shearing head as appropriate, see
Chapter Lubricating the shearing head.
Check the product for any damage, abnormal run-
ning noise, overheating or excessive vibration.
Inspect the shearing head for broken, damaged or
blunt cutting edges.
Set aside the product if it is damaged.