Technical data for the OMRON Comp Air compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebuliser kit:
Particle Size:
Medication Tank
Capacity:
Appropriate Medication
Quantities:
Sound:
Nebulisation Rate:
Aerosol Output:
Aerosol Output Rate:
Notes:
• Performance may vary with drugs such as suspensions or high viscosity. See
drug supplier's data sheet for further details.
• The data above is referenced by Omron Healthcare.
• See web site of Omron Healthcare Europe to update technical information.
Result of cascade impactor
Air (NE-C28-E) and OMRON V.V.T. Nebuliser kit
100
90
80
70
60
50
40
30
20
10
0
0.1
*MMAD = Mass Median Aerodynamic Diameter
**Independently measured at SolAero Ltd., Canada, Dr. John Dennis, according to
EN 13544-1
*MMAD **approximately 3µm
7 ml maximum
2 ml minimum - 7 maximum
Noise level (at 1 m distance) 60 db
0.4 ml/min (by weight loss)
**0.4 ml (2ml, 1%NaF)
**0.06 ml/min (2ml, 1%NaF)
measurements for particle size with OMRON Comp
**
Cumulative % particle mass of sodium fluoride under size
1
Particle Size Dp (µm)
9. Technical data
Individual tests
Mean
MMAD: 3µm
10
100