8. Instrument messages during operation
Display
Description
During the acquisition of a calibration point, an invalid value is read from the transdu-
Er.Pnt
cer.
Is displayed if, during the selection of the function "NAStEr
Err.rF
EN
the power-saving mode of the indicator, a SLAVE instrument is not found after 60
seconds. This message is displayed if the radio module is activated.This message is
displayed for 3 seconds and then the display is switched off.
Calibration error: Too low a comparison weight has been used; it is recommended to
Er
11
use a weight that corresponds to at least half the nominal load of the scales.
Calibration error: The acquired calibration point ("tP1
Er
12
the zero point ("tP0
The number of points of the transducer for the subdivision of the scales is less than
Er
37
two. Carry out the calibration again (correctly), paying attention to the nominal load
and the scale divisions.
Occurs when the instrument is not yet initialised and calibrated.
Er
39
When the instrument displays
the SETUP area. Carry out the initialisation of the display unit (Parameter
the selection of the keypad type (parameter
ters of the SETUP area and the calibration of the instrument.
Occurs when the instrument has been initialised, but not yet calibrated.
ER
85
When the instrument displays
the SETUP area.
During calibration, internal, negative points have been calculated:
C.Er.36
(blinking)
undEr
The scale is in underload (i.e. the weight on the scale is ≤ 100 scale divisions under
the gross zero, when the instrument is legally calibrated). This message is accompa-
nied by an acoustic signal.
(blinking)
oVEr
The scales are in overload (i.e. the weight on the scales is 9 scale divisions greater
than the nominal load/weighing capacity). This message is accompanied by an
acoustic signal.
Is shown for a moment when the master instrument manages to connect to the slave
Eco H
instrument X. When the connection is not possible, the message persists and the
master instrument emits an acoustic signal.
The connected tilt sensor is preventing a weighing process.
tilt
The message is displayed when the tilt of a pallet truck is greater than 2 % or, with a
fork-lift truck, is greater than 5 %.The activation of the "TILT alarm
approx. 3 seconds delay.
34
"
).
The calibration point lies below the zero point.
The signal is negative (check the cable connections to the load cell).
WIKA operating instructions SF
"
"
"
, press the TARE button, in order to access
ER
39
"
"
), the programming of all parame-
keyb
"
"
, press the TARE button, in order to access
ER
85
"
and the activation of
"
"
or "tP2
or "tP3
) is the same as
"
dEFAu
"
is carried out with
gas filling cart, model GFU08
6
"
),