4 Technical description
run
view
STOP
10
RUN
Starts the run and puts the program into RUN mode. The light (HV on) is switched on.
VIEW
In run mode, VIEW switches between the actual values for voltage, current and elapsed
time. Pressing VIEW for more than three seconds starts automatic switching between the
actual values for voltage, current and elapsed time. This automatic switching is stopped
by pressing VIEW or RUN.
In End mode, VIEW switches between the end parameter values.
STOP
Stops the run and puts the instrument in End mode. The voltage and the light (HV on) are
switched off. By pressing VIEW the end parameter values are displayed.
4.5

Output sockets

There are two sets of output sockets connected in parallel to allow two electrophoresis
units to be connected and run at the same time, see fig. 1 inside front cover.
4.6

Rear panel

The rear panel is shown in fig 2, see inside front cover. On the rear panel there is:
1.
A mains switch. Press in I to switch on the power to the power supply. Press 0 to
switch off the power.
2.
A socket for the mains cable.
3.
A switch for voltage range. The left position corresponds to 100–120 V and the right
to 220–240 V.
4.
Vents.
Electrophoresis Power Supply EPS 301 User Manual 18-1130-18 Edition AC