8 Technical specifications

8 Technical specifications
16
Regulation
Output mode
Maximum Power Output
Programming range
Output resolution
Programming resolution
Accuracy
Line regulation
Load regulation
Ripple
Short term stability
Long term stability
Output protection
Recovery after power failure
Ambient operating temperature
Ambient operating humidity
Ambient operating pressure
Mains requirements
Power consumption
Dimensions (WxDxH)
Weight
Electrophoresis Power Supply EPS 301 User Manual 18-1130-18 Edition AC
Maximum voltage and current with automatic cross-over
at preset limits
Contante Voltage: 0–300 V DC,
Contante Current: 0–400 mA
80 W
Voltage: 5–300 V DC
Current: 10–400 mA
Time: 00.01–23.59 h or off (- - - -)
Voltage: 1 V
Current: 1 mA
Voltage: 1 V
Current: 1 mA
Time: 1 min
Voltage: 4%, ± 2 V
Current: 4%, ± 4 mA
Timer: 0.1% ± 1 min
< 0.5%
< 1% at load change 10–90% of maximum load
< 3% at 300 V
< 0.2% /10 h after warm up
< 1% /year
Fully protected against any overload conditions
The program continues automatically
4–40 °C
0–95%
68–106 kPa, maximum altitude of 2000 m
100–120 V/220–240 V; 50/60 Hz
Max 120 W
250 x 215 x 95 mm
3.0 kg