TRACK BELT REPLACEMENT (FOR MODEL 6252) - CONTINUED
4.
Using a 5/32" hex key, remove the two button head cap
screws from the track spacer pivots and remove the pivots.
5.
Using two 7/16" wrenches, remove the hex head cap screw, nut and
washers from both head end track rollers and remove the rollers.
6.
Remove the belts by threading them through the wheel
support plate on the side where the parts were removed.
www.stryker.com
Service Information
6250-001-161 REV D
Figure 49 - Remove the track spacer pivots
Figure 50 - Remove the rollers
Figure 51 - Remove the worn belts