3.
A1
4.
When A2 and A3 are more than 0.35 mm / m apart, an adjustment
!
is necessary. Contact your authorised dealer or else the UMAREX-
LASERLINER Service Department.
Checking the vertical line
Position the device about 5 m from a wall. Fix a plumb bob with a line of
2.5 m length on the wall, making sure that the bob can swing freely. Switch
on the device and align the vertical laser to the plumb line. The precision is
within the specified tolerance if the deviation between the laser line and
the plumb line is not greater than ± 1.75 mm.
Checking the horizontal line
Position the device about 5 m from a wall and
switch on the cross laser. Mark point B on the
wall. Turn the laser cross approx. 2.5 m to the
right and mark point C. Check whether the
horizontal line from point C is level with point B
to within ± 1.75 mm. Repeat the process by turning
the laser to the left.
!
Regularly check the calibration before use, after transport and after
extended periods of storage.
Calibration
The meter needs to be calibrated and tested on a regular basis to ensure
it produces accurate measurement results. We recommend carrying out
calibration once a year.
14
EN
4. Turn the device
A2
A2
A3
A2
A3
through 180° and
mark point A3. The
difference between
points A2 and A3 is
the tolerance.
2.5 m
B
C